Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
310
ISQ 7000 Hardware Manual
Thermo Scientific
20-Pin Feedthrough
To replace the 20-pin feedthrough:
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel and left side panel screws
from the interior front panel. Then slide the left and top cover panels toward the back of
the instrument and lift them off.
3. The lens driver board is the first board standing on end in the center of the instrument.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or the column. However, you
must vent the instrument.
Figure 334.
Locating the Lens Driver Board
Lens Driver Board
Bracket
Top Manifold
Cover