3
Advanced Maintenance
Replacing Components of the Electron Multiplier
Thermo Scientific
ISQ 7000 Hardware Manual
265
Anode Feedthrough
To replace the anode feedthrough, o-ring, four screws, and a standoff
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel screw from the interior
front panel. Then slide the top cover panel toward the back of the instrument and lift it
off.
3. Disconnect the electrometer board cable from the electron multiplier.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or the column. However, you
must vent the instrument.
Figure 281.
Disconnecting the Electrometer Board Cable from the Electron Multiplier
Electron Multiplier
Plate
Electrometer
Board Cable
Electrometer
Shield