3
Advanced Maintenance
Replacing Components of the Ion Source
Thermo Scientific
ISQ 7000 Hardware Manual
217
To replace components of the ExtractaBrite ion source cartridge
1. Follow the procedure in
“Removing the Ion Source Cartridge”
to remove the
entire ion source cartridge.
2. Replace the components as necessary.
3. Reassemble and reinsert the ion source cartridge by following the steps in
.
To replace components of the Advanced EI ion source cartridge
1. Follow the procedure in
“Removing the Advanced Ion Source Cartridge on a ISQ 7000
to remove the ion source cartridge.
2. Replace the components as necessary
3. Reassemble and reinsert the ion source cartridge by following the steps in
the Advanced EI Ion Source Cartridge”
.
Figure 208.
Replaceable Components of the Advanced EI Ion Source Cartridge
AEI Filament
Shield
Lens 3/RF
Lens
AEI Ion Volume
Insulator
Lens 2
AEI Ion Cartridge
Sleeve
AEI Ion Volume
AEI Filament
Insulator
Filament
Vespel Cap
AEI Filament
Spacer