3
Advanced Maintenance
Replacing Components of the Analyzer
160
ISQ 7000 Hardware Manual
Thermo Scientific
5. Use a T20 Torxhead screwdriver to remove the two screws on the back of the ion source.
6. Remove the ion source from the analyzer tray and set it aside.
7. Replace the analyzer tray with a new one if needed.
Figure 128.
Reattaching the Ion Source to the Analyzer Tray
Figure 129.
Replacing the Analyzer Tray
Screws
Ion Source
Block
Grounding
Strap