Advanced Maintenance
Replacing Components of the Manifold
Thermo Scientific
ISQ 7000 Hardware Manual
247
11. Flip the manifold door over and remove the o-ring.
12. Flip the door over and twist the vent valve knob and o-ring to remove them.
13. Replace the front manifold plate.
14. Insert and tighten the vent valve knob and o-ring into the front manifold plate. Be sure
not to pinch the o-ring.
15. Reattach the front plate o-ring.
• Flip the manifold door over and reattach the magnet yoke, source magnets, and
source retainer magnets.
16. Reattach the three screws and valve body to the front manifold plate.
Figure 257.
Removing the Manifold O-Ring
Figure 258.
Removing the Vent Valve Knob and O-Ring
Figure 259.
Replacing the Front Manifold Plate
Manifold
O-ring
Vent
Valve
Knob
Vent Valve
O-ring