Advanced Maintenance
Replacing Components of the Analyzer
Thermo Scientific
ISQ 7000 Hardware Manual
175
Quadrupole Wire Set
To replace the quadrupole wire set:
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Remove the analyzer tray from the instrument by following the instructions in
“Removing the Analyzer Tray from the ISQ 7000 Instrument”
3. Remove the screws from each side of the exit endcap.
4. Remove the quadrupole rods and exit endcap from the tray.
5. Detach the exit endcap from the quadrupole rods.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Figure 147.
Removing the Quadrupole and Ion Guide from the Analyzer Tray
Screws
Exit Endcap