Advanced Maintenance
Replacing the Power Supplies
350
ISQ 7000 Hardware Manual
Thermo Scientific
4. Disconnect the large feedthrough cables.
5. Use a T20 Torxhead screwdriver to remove the screws connecting the dynode/multiplier
power supply to the back and bottom of the chassis.
Figure 390.
Disconnecting the Large Feedthrough Cables
Figure 391.
Removing the Dynode/Multiplier Power Supply
Electron Multiplier
Plate
Large
Bottom
Screw
Back
Screws
Dynode/Multiplier
Power Supply