1
Performing Routine Maintenance
Replacing a Column
Thermo Scientific
ISQ 7000 Hardware Manual
17
3. On the left side of the instrument, push down on the power switch to power-off the
ISQ 7000 system.
.
4. Open the front door of the instrument.
5. Open the front door of the instrument.
6. Look behind the right side of the vacuum interlock shield and twist the vent valve knob
one and a half times in a counterclockwise direction to open the vent.
7. Remove the current column from the MS transfer line. To replace the column in the GC
inlet, follow the instructions in
“Installing the Column in the GC Inlet”
8. Connect the column to the MS transfer line. When connecting the column to the
transfer line, you may use either the spring loaded transfer line nut with the graphite
Vespel™ ferrule or the regular transfer line nut
.
Figure 6.
Powering Off the ISQ 7000 Instrument
Note
For best results, we recommend you use the spring loaded transfer line nut.
Power Switch
Vacuum Interlock Shield
Vent Valve
Knob
Hydrogen Safety Screws
Hydrogen Safety Screw
(On Front Door Panel)