Advanced Maintenance
Replacing Components of the Analyzer
154
ISQ 7000 Hardware Manual
Thermo Scientific
8. Disconnect the wires that connect the quadrupole and ion guide to the 4-pin feedthrough
in the vacuum manifold.
9. Open the manifold door and slide the filament board and wires off the pins on the source
interface board.
10. Remove the column and replace with a nut and no-hole ferrule.
11. Pull the GC away from the mass spectrometer.
12. Remove the instrument right-side panel to access the transfer line. See
.
Figure 120.
Removing the Right Side Panel
13. Identify the set screw on the source heater block. This T10 screw holds the transfer line in
alignment with the ion volume heater block. See
.
Figure 119.
Disconnecting the 4-Pin Connector Wires
Wires
4-pin
Connector