3
Advanced Maintenance
Replacing Components of the Manifold
248
ISQ 7000 Hardware Manual
Thermo Scientific
17. Reattach the two screws and pillow block to the front manifold plate.
18. Hold the spring and safety interlock in place as you attach the vacuum interlock shield
and screws.
19. Reconnect the vacuum interlock cable and tubing.
20. Restart your system by following the instructions in