Performing Routine Maintenance
Cleaning the Ion Source Cartridge
64
ISQ 7000 Hardware Manual
Thermo Scientific
15. Set the components on a clean work surface.
Removing the Advanced Ion Source Cartridge on a ISQ 7000 System
1. Click
Shut Down
on the TSQ Series Dashboard.
2. Click
Yes
to continue the shutdown process. The high voltages, heaters, and
turbomolecular pump power off. Once the turbomolecular pump reaches 50% speed, or
five minutes elapses, the foreline pump powers off and you may vent the system.
3. Reach around the left side to the back of the instrument and push down on the power
switch to power-off the instrument.
4. Open the front door of the ISQ 7000 mass spectrometer.
Figure 53.
Ion Source Cartridge Components
Ion Volume
Locking Ring
Repeller
Lens 1
Lens 2
Lens 3/RF Lens
Ion Cartridge Sleeve
EI Ion Volume
Repeller
Spring
Repeller
Nut
Ion
Volume-Repeller
Insulator
Note
You must shut down and vent the mass spectrometer even if your instrument has a
vacuum interlock.
Note
The amber vacuum light on the front of the instrument starts blinking rapidly,
indicating the mechanical pump has powered off after a five minute period with the
turbomolecular pump off (such as when the instrument is shut down), or due to a
sustained vacuum fault lasting five minutes. When the turbomolecular pump spins down
below 50% speed due to the shut down process, the vacuum light turns off.