3
Advanced Maintenance
Replacing Components of the Ion Source
Thermo Scientific
ISQ 7000 Hardware Manual
211
Lens Plate and Springs
To replace the lens plate and springs
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Remove the analyzer tray from the instrument by following the instructions in
“Removing the Analyzer Tray from the ISQ 7000 Instrument”
3. Disconnect the lens wires form the source interface board with a slotted screwdriver.
4. Use a T10 Torxhead screwdriver to disconnect the lens plate and springs from the lens
heater block.
5. Pull the upper portion of the lens plate out so the contacts are free of the lens heater
block. Slide the lens plate and springs down to clear it from the lens heater block.
6. Attach a new lens plate and springs to the lens heater block and reattach the screws.
7. Connect the lens wires form the source interface board with a slotted screwdriver.
8. Once the components are correctly attached to the analyzer tray, follow the instructions
in
9. Restart your system by following the instructions in
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering
these components.
Figure 199.
Replacing the Lens Plate and Springs
Lens Plate and
Springs
Screw
Lens Heater Block