3
Advanced Maintenance
Replacing Components of the Analyzer
Thermo Scientific
ISQ 7000 Hardware Manual
181
Exit Endcap
To replace the exit endcap
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Remove the analyzer tray from the instrument by following the instructions in
“Removing the Analyzer Tray from the ISQ 7000 Instrument”
3. Remove the two screws on each side of the exit endcap and slide it off the back of the tray.
4. Attach a new exit endcap and reattach the two screws.
5. Follow the instructions in
6. Restart your system by following the instructions in
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Figure 155.
Removing the Exit Endcap from the Analyzer Tray
Tip
You will get better signal if you gently push the exit endcap forward while tightening
the screws.
Screws
Exit Endcap