Performing Routine Maintenance
Cleaning the Ion Source Cartridge
Thermo Scientific
ISQ 7000 Hardware Manual
59
21. Let the ion source cartridge cool down before removing the components from the source
holder.
22. Disassemble the ion source cartridge by removing the locking ring first, then the repeller
spring, then the nut, insulator, and repeller (which comes out in one piece), ion volume,
lens 1, lens 2, and lens 3/RF lens.
23. Set the components on a clean work surface.
CAUTION BURN HAZARD:
The ion source cartridge may be hot, so use caution.
ATTENTION RISQUE DE BRÛLURE
: soyez prudent, car la cartouche de la source d’ions
peut être chaude.
IMPORTANT
Always wear clean, lint-free gloves when handling the ion source cartridge.
Figure 46.
Ion Source Cartridge Components
Ion Volume
Locking Ring
Repeller
Lens 1
Lens 2
Lens 3/RF Lens
Ion Cartridge Sleeve
EI Ion Volume
Repeller
Spring
Repeller
Nut
Ion
Volume-Repeller
Insulator