Replacing Components of the Vacuum Interlock
Thermo Scientific
ISQ 7000 Hardware Manual
233
29. Reattach the three screws holding the valve body to the front manifold plate.
30. Reattach the two screws connecting the pillow block to the manifold door.
Figure 236.
Reattaching the Valve Body to the Front Manifold Plate
Figure 237.
Reattaching the Pillow Block
Screws
Front
Manifold
Plate
Pillow
Block
Screws
CAUTION INSTRUMENT DAMAGE:
If the vacuum interlock handle is not fully down
and the ball valve is not fully closed, then the ball valve may not be able to completely
open it during use and it could damage the ball valve and probes.