Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
Thermo Scientific
ISQ 7000 Hardware Manual
295
9. Pry the 7-pin connector on the rod driver board away from the RF board.
10. If necessary, replace the rod driver board.
11. Attach the rod driver board to the new RF board.
Figure 312.
Disconnecting the Rod Driver Board 7-Pin Connector
Figure 313.
Replacing the Rod Driver Board
7-Pin Connector