3
Advanced Maintenance
Replacing Components of the Electron Multiplier
Thermo Scientific
ISQ 7000 Hardware Manual
267
7. Remove the electrometer board to reveal the anode feedthrough.
8. Use a T10 Torxhead screwdriver to remove the two screws and 5.5 mm nut driver to
remove the standoff located on the outer rim of the anode feedthrough, but leave the
center pin of the standoff intact. The standoff serves as the ground for the electrometer
board. The center pin is the signal path for the electrons generated in the electron
multiplier going into the electrometer.
Figure 284.
Removing the Anode Feedthrough, O-ring, Standoff and Screws
Anode Feedthrough
and O-ring
Standoff
Screws