3
Advanced Maintenance
Replacing Components of the Ion Source
196
ISQ 7000 Hardware Manual
Thermo Scientific
Thumbscrew-Repeller Spacers
To replace the thumbscrew-repeller spacers
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
3. Open the manifold door and twist off the source-repeller thumbscrews one at a time to
prevent the repeller plate from falling out.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Figure 179.
Opening the Manifold Door
Figure 180.
Removing the Source-Repeller Thumbscrews
Screws
Source-Repeller
Thumbscrews
Содержание ISQ 7000
Страница 1: ...ISQ 7000 Mass Spectrometers Hardware Manual 1R120617 0002 Revision C December 2018 ...
Страница 8: ......
Страница 174: ...2 Troubleshooting Contacting Technical Support 146 ISQ 7000 Hardware Manual Thermo Scientific ...
Страница 400: ...3 Advanced Maintenance Replacing Fans and Filters 372 ISQ 7000 Hardware Manual Thermo Scientific ...