Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
Thermo Scientific
ISQ 7000 Hardware Manual
337
4. Disconnect the electrometer board cable and large feedthrough cable.
5. The RF board is the second board standing on end toward the center of the ISQ 7000
instrument. Use a T20 Torxhead screwdriver to remove the screw holding the board and
bracket to the chassis.
Figure 372.
Disconnecting the Electron Multiplier Cables
Figure 373.
Removing the RF Board and Bracket
Electron Multiplier
Plate
Large
Electrometer
Board Cable
RF Board
Chassis
Screw
Bracket