1
Performing Routine Maintenance
Cleaning the Ion Source Cartridge
Thermo Scientific
ISQ 7000 Hardware Manual
63
Figure 52.
Removing the Ion Source Cartridge
13. Let the ion source cartridge cool down before removing the components from the source
holder.
14. Disassemble the ion source cartridge by removing the locking ring first, then the repeller
spring, then the nut, insulator, and repeller (which comes out in one piece), ion volume,
lens 1, lens 2, and lens 3/RF lens.
CAUTION - BURN HAZARD
The ion source cartridge may be hot, so use caution.
ATTENTION RISQUE DE BRÛLURE
: soyez prudent, car la cartouche de la source d’ions
peut être chaude.
IMPORTANT
Always wear clean, lint-free gloves when handling the ion source cartridge.
Содержание ISQ 7000
Страница 1: ...ISQ 7000 Mass Spectrometers Hardware Manual 1R120617 0002 Revision C December 2018 ...
Страница 8: ......
Страница 174: ...2 Troubleshooting Contacting Technical Support 146 ISQ 7000 Hardware Manual Thermo Scientific ...
Страница 400: ...3 Advanced Maintenance Replacing Fans and Filters 372 ISQ 7000 Hardware Manual Thermo Scientific ...