3
Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
Thermo Scientific
ISQ 7000 Hardware Manual
297
16. Reattach the top cover panel and tighten the screw holding it in place.
17. Restart your system by following the instructions in
Note
The new RF board will have a slightly different performance than your current
board. It will need to be adjusted by a Field Service Engineer because the adjustments
must be made while the system is running.