4
Upgrade Equipment
Turbomolecular Pumps
Thermo Scientific
ISQ 7000 Hardware Manual
405
22. Connect the turbo pump cable to the 15-pin connector on the distribution board.
23. Replace the right side panel and attach the screws holding it in place.
24. Slide the instrument closer to the GC.
25. Reattach the top cover panel and tighten the screw holding it in place.
26. Restart your system by following the instructions in
Figure 465.
Reconnecting the Turbo Pump Cable to the Distribution Board
Turbo Pump
Cable