1
Performing Routine Maintenance
Maintaining the Foreline Pump
Thermo Scientific
ISQ 7000 Hardware Manual
41
3. Click
Shut Down
on the ISQ Dashboard or on the ISQ 7000 status page within the
instrument software.
4. Click
Yes
to continue the shutdown process. The high voltages, heaters, and
turbomolecular pump power off. Once the turbomolecular pump reaches 50% speed, or
five minutes elapses, the foreline pump powers off and you may vent the system.
5. Reach around the right side to the back of the instrument and push down on the power
switch to power-off the ISQ 7000 instrument.
6. If you are using hydrogen as a carrier gas, unscrew the hydrogen safety screw on the front
door.
.
7. Open the front door of the instrument.
8. Look behind the right side of the vacuum interlock shield and twist the vent valve knob
one and a half times in a counter-clockwise direction to open the vent.
Note
The amber vacuum light on the front of the instrument starts blinking rapidly,
indicating the mechanical pump has powered off after a five minute period with the
turbomolecular pump off (such as when the instrument is shut down), or due to a
sustained vacuum fault lasting five minutes. When the turbomolecular pump spins down
below 50% speed due to the shut down process, the vacuum light turns off.
Figure 29.
Powering Off the ISQ 7000 Instrument
Power Switch
Vacuum Interlock Shield
Vent Valve
Knob
Hydrogen Safety Screws
Hydrogen Safety Screw
(On Front Door Panel)