3
Advanced Maintenance
Replacing Components of the Analyzer
168
ISQ 7000 Hardware Manual
Thermo Scientific
8. Look inside the endcap and use a T10 Torxhead screwdriver to remove the two screws
holding the ion guide to the ion guide clamp.
9. Rotate the ion guide clamp until the ion guide can be removed.
10. Insert the new ion guide into the ion guide clamp and entrance endcap.
11. Rotate the ion guide clamp until the ion guide is locked in place.
12. Slide the ion guide and entrance endcap back onto the rods and exit endcap.
13. Reattach the ion guide and quadrupole to the analyzer tray. Make sure the quad and ion
guide and fully forward before tightening screws.
Figure 137.
Detaching the Ion Guide from the Ion Guide Clamp
Figure 138.
Replacing the Ion Guide
Screws
Ion Guide Clamp
Ion Guide