3
Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
300
ISQ 7000 Hardware Manual
Thermo Scientific
7. Push the board down into the connectors.
8. Gently push the board toward the instrument, carefully aligning the board over the
feedthrough pins, until it stops.
9. Reattach the screw holding the RF board to the bracket.
10. Reattach the top cover panel and tighten the screw holding it in place.
11. Restart your system by following the instructions in
Figure 319.
Reattaching the RF Board and Bracket
Note
This screw is required for good ground contact.
Note
The new RF board will have a slightly different performance than your current
board. It will need to be adjusted by a Field Service Engineer because the adjustments
must be made while the system is running.
RF Board
Chassis
Screw
Bracket