Performing Routine Maintenance
Configuring the ISQ 7000 Mass Spectrometer
6
ISQ 7000 Hardware Manual
Thermo Scientific
Configuring the ISQ 7000 Mass Spectrometer
When you receive your ISQ 7000 mass spectrometer, a Field Service Engineer will configure
the system for you. However, if you need to reconfigure your ISQ system, follow these steps:
1. Place the ISQ 7000 instrument, GC, computer, monitor, and keyboard on a work table
and the foreline pump on the floor underneath.
2. Connect all the cables to the computer, keyboard, mouse, monitor, and printer (if
applicable).
3. Connect the handshaking cable to the GC and to the GC Start connector on the ISQ
7000 instrument.
4. Connect the computer and monitor electrical cables to the wall outlet.
Note
If you are using CI, connect the reagent gas tubing to one of the Reagent Gas
connectors on the ISQ 7000 instrument. See
for details.
Figure 1.
Configuring the ISQ 7000 Mass Spectrometer
Reagent
Gas (A & B)
Connectors
GC Start
Connector
Power Switch
Data
System
Connector
System or
Optional
Mech Pump
Connections
AC Input
Connector
Accessory Start
Connector