Troubleshooting
Investigating Filament and Lens Control Issues
Thermo Scientific
ISQ 7000 Hardware Manual
129
Check the source interface board to make sure the filament wires are fully connected. See
Source Interface Board and Heat Shield
for
details.
Make certain there are no large leaks. High pressure readings disable the lens voltages.
Make certain the lever on the vacuum interlock is fully down. The lenses are disabled when the lever is in the up position.
Possible Remedies