3
Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
312
ISQ 7000 Hardware Manual
Thermo Scientific
8. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
9. Detach the filament board from the source interface board.
10. Open the manifold door and remove the bottom left screw holding the metal heat shield
to the source interface board.
Figure 336.
Opening the Manifold Door
Figure 337.
Removing the Heat Shield
Screws
Source Interface Board
Heat Shield
Screw
Filament Board