3
Advanced Maintenance
Replacing Components of the Analyzer
Thermo Scientific
ISQ 7000 Hardware Manual
159
Removing the Analyzer Tray Components
To remove the analyzer tray components from the tray
1. Remove the wire connecting the quadrupole entrance lens to the source interface board
by unscrewing the terminal with a slotted screwdriver.
2. Remove the screws on each side of the endcaps.
3. Remove the entire quadrupole, as well as the ion guide, from the analyzer tray. Set them
all aside.
4. Use a T10 Torxhead screwdriver to remove the lens heater block grounding strap screw
from the analyzer tray.
Figure 127.
Removing the Quadrupole and Ion Guide from the Analyzer Tray
Note
Make sure you set the rods on a clean, lint-free cloth. Dust and lint on the rods lead
to poor analytical performance.
Screws
Exit Endcap
Entrance Endcap