Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
Thermo Scientific
ISQ 7000 Hardware Manual
335
6. Set the analyzer tray on a clean, lint-free surface.
7. Hold the new source interface board next to the old board.
8. As you remove each of the wires from the old board, connect them to the same connectors
on the new board.
9. Follow the instructions in
to reinstall the
analyzer tray.
10. Restart your system by following the instructions in
Figure 370.
Reconnecting the Source Interface Board Wires
Source Interface
Board Wires