3
Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
Thermo Scientific
ISQ 7000 Hardware Manual
315
15. Look through the top of the instrument and align the pins on the 20-pin feedthrough
with the holes in the source interface board to connect them together.
16. Reattach the filament board to the source interface board.
17. Slide the heat shield into its slot and reattach the screw holding it in place.
Figure 342.
Reattaching the Source Interface Board to the 20-Pin Feedthrough
Figure 343.
Reattaching the Heat Shield
20-Pin
Feedthrough
20-Pin Feedthrough
Pins in Source
Interface Board
Source Interface
Board
Heat Shield
Screw
Filament Board