Instruction Manual
HI 93732N
Dissolved Oxygen ISM
w w w . h a n n a i n s t . c o m
This Instrument is in
Compliance with the CE Directives
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
WARRANTY
WARRANTY
WARRANTY
WARRANTY
WARRANTY
HI 93732N is warranted for two years
against defects in
workmanship and materials when used for its intended purpose and
maintained according to instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of purchase,
serial number and the nature of the failure. If the repair is not covered
by the warranty, you will be notified of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service
department and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete
protection.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments
Inc., Woonsocket, Rhode Island, 02895 , USA.
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at [email protected].
This instrument is in compliance with directives.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• Two Sample Cuvets and Caps
• One Transport Cap
• One 60 mL glass Bottle with Stopper
• 9V Battery
Note:
Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing with
the supplied accessories.
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
GENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTION
The
HI 93732N
meter measures the content of dissolved
oxygen (O
2
) in
surface, feed, natural and waste waters in
the 0.0 to 10.0 mg/L (ppm) range.
The meter uses an exclusive positive-locking system to ensure
that the cuvet is in the same place every time it is placed
into the measurement cell.
The reagents are in liquid form and are supplied in bottles.
The amount of reagent is precisely dosed to ensure maximum
repeatability.
Display codes aid the user in routine operations. The meter
has an auto-shut off feature that will turn itself off after 10
minutes of non-use.
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Range
0.0 to 10.0 mg/L (ppm)
Resolution
0.1 mg/L
Accuracy
±0.2 mg/L ± 3% of reading
Typical EMC
±0.1 mg/L
Deviation
Light Source
Light Emitting Diode @ 470 nm
Method
Adaptation of the
Standard Methods for
the Examination of Water and Wastewater
(18
th
edition),
azide modified Winkler
method. The reaction between dissolved
oxygen and the reagent causes a yellow
tint in the sample.
Light Life
Life of the instrument
Light Detector
Silicon Photocell
Environment
0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery
1 x 9 volt/40 hours
Type/Life
Auto-Shut off
After 10' of non-use
Dimensions
180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight
290 g (10 oz.).
REQUIRED REAGENTS
Code
Description
Quantity
HI 93732
A
-0
Reagent A
5 drops
HI 93732
B
-0
Reagent B
5 drops
HI 93732
C
-0
Reagent C
10 drops
REAGENT SETS
HI 93732-01
Reagents for 100 tests
HI 93732-03
Reagents for 300 tests
This indicates that the meter is in a
ready state and zeroing can be performed.
Sampling in Progress. This prompt
appears each time the meter is performing
a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Insert a
sample before adding reagent and press
ZERO.
Under range. A blinking "0.00" indicates
that the sample absorbs less light than
the zero reference. Check the procedure
and make sure you use the same cuvet
for reference (zero) and measurement.
Over range. A flashing value higher than
the maximum concentration readable (see
specifications) indicates that the sample
absorbs too much light, meaning that
the concentration is too high.
Light over range. The cuvet is not inserted
correctly and an excess ambient light is
reaching the detector. If the cover is
properly installed, then contact your dealer
or the nearest Hanna Customer Service
Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is not
the case, contact your dealer or the
nearest Hanna Customer Service Center.
The "V" indicates that the battery voltage
is getting low and the battery needs to
be replaced.
This indicates that the battery is dead
and must be replaced.
Note:
once this indication is displayed,
the meter will lockup. Change the
battery to restart.
DISPLAY CODE GUIDE