3
Advanced Maintenance
Replacing Components of the Ion Source
Thermo Scientific
ISQ 7000 Hardware Manual
201
Repeller Plate
To replace the repeller plate:
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Remove the analyzer tray from the instrument by following the instructions in
“Removing the Analyzer Tray from the ISQ 7000 Instrument”
3. Twist off the source-repeller thumbscrews.
4. Remove the thumbscrew-repeller spacers.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Figure 186.
Removing the Source-Repeller Thumbscrews
Figure 187.
Removing the Thumbscrew-Repeller Spacers
Source-Repeller
Thumbscrews
Thumbscrew-Repeller
Spacers