Advanced Maintenance
Replacing Components of the Ion Source
202
ISQ 7000 Hardware Manual
Thermo Scientific
5. Remove the insulating spacers.
6. Remove the repeller plate.
7. Disconnect the repeller plate wire from the source interface board.
8. Connect the new repeller plate wire to the source interface board.
9. Attach the new repeller plate.
10. Reattach the insulating spacers.
11. Reattach the thumbscrew-repeller spacers.
12. Reattach the source-repeller thumbscrews
Figure 188.
Removing the Insulating Spacers
Figure 189.
Removing the Repeller Plate
Note
When reattaching the spacers, make sure they fit into the grooves on the
thumbscrews. Also, do not over-tighten the thumbscrews or it may crush the spacers.
Insulating Spacers
Repeller Plate
Wire