3
Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
Thermo Scientific
ISQ 7000 Hardware Manual
329
21. Slide the main power supply back down into its slot. Once it is in the slot, squeeze the
sides of the instrument together slightly while you are reattaching the screws on the back
of the instrument.Otherwise the top cover panel will not fit perfectly.
22. Reattach the dynode/multiplier power supply to the back and bottom of the chassis by
reattaching the three screws.
Figure 360.
Reinstalling the Power Supply System
Figure 361.
Reinstalling the Dynode/Multiplier Power Supply
Power Supply System
Screws
Bracket
Screw
Screws
Connector