Performing Routine Maintenance
Cleaning the Ion Source Cartridge
54
ISQ 7000 Hardware Manual
Thermo Scientific
9. Slowly push the source exchange tool into the instrument. Because you are removing the
ion volume in this step, you will go down the right side of the metal track on the barrel.
10. Once the tool is all the way in (when the groove at the end of the barrel is covered by the
black handle) twist the handle to the left to engage the ion source cartridge and move it
onto the end of the barrel. You are essentially disconnecting the ion source cartridge from
the ion source block and moving it onto the tool.
Figure 37.
Pushing the Tool into the Instrument