3
Advanced Maintenance
Replacing Components of the Front Door
Thermo Scientific
ISQ 7000 Hardware Manual
287
Front Door Latch
To replace the latch on the front door
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the left side panel screw from the interior front
panel.
3. Slide the left side panel toward the back of the instrument and lift it off.
4. Use a T10 Torxhead screwdriver to remove the two screws holding the latch onto the
front door.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or the column. However, you
must vent the instrument.
Figure 302.
Removing the Latch from the Front Door
Screws
Latch