3
Advanced Maintenance
Replacing Components of the Vacuum Interlock
Thermo Scientific
ISQ 7000 Hardware Manual
225
6. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
7. Set the manifold door on a clean work surface and remove the two screws holding the
pillow block in place.
Figure 222.
Detaching the Manifold Door
Figure 223.
Removing the Pillow Block
Screws
Pillow Block
Screws