Performing Routine Maintenance
Cleaning the Ion Source Cartridge
66
ISQ 7000 Hardware Manual
Thermo Scientific
8. Open the manifold door.
9. Let the system cool down for 30 minutes.
10. Disconnect the filament wires from the source interface board.
Figure 56.
Disconnect the Filament Wires
11. Grasp the Vespel cap and gently remove the Advanced EI ion source from the ion source
block. See
Figure 55.
Removing the Manifold Door Screws
CAUTION - BURN HAZARD
The ion source cartridge may be hot, so use caution.
ATTENTION RISQUE DE BRÛLURE
: soyez prudent, car la cartouche de la source d’ions
peut être chaude.
Filament Wires