1
Performing Routine Maintenance
Replacing the Electron Multiplier
Thermo Scientific
ISQ 7000 Hardware Manual
107
35. Allow sufficient time for the vacuum and temperatures to stabilize.
36. Go back to the ISQ Dashboard and click
Auto Tune
to open
ISQ 7000 Autotune
.
37. Run any tune that sets the gain of the electron multiplier to 3 x 10
5
.