Advanced Maintenance
Upgrading the ISQ 7000 System Software
Thermo Scientific
ISQ 7000 Hardware Manual
273
6. Click on the
NIST Browser
button to install the NIST MS Browser. This functionality is
required to allow automatic searching of libraries from Xcalibur or layered applications.
Install the NIST Browser according to the installation instructions. The NIST installer
will automatically create the required folders for the browser and demo libraries. For best
operation, do not change the default NIST Browser installation locations.
7. This completes the Xcalibur Core Components installation for GC-MS. To install the
GC and MS drivers that are needed for instrument operation, refer to the installation
instructions that are provided with those discs.
8. Click
Exit
to leave the Xcalibur 4.0 Setup window.
Configuring LAN Adapter Inactivity Time
When the TRACE GC Ultra is equipped with a LAN CPU card, the inactivity timeout
setting needs to be set to 30,000 ms to effectively maintain communication with the software,
even in periods of heavy data processing. Depending on the serial number of your TRACE
GC Ultra, the software tool used to manage the LAN CPU may vary.
To configure the NPort Management Suite
1. Open the NPort Management Suite by selecting
Start | All Programs | NPort
Management Suite | Configurator
.
2. Click the magnifying glass button on the tool bar or click
Locate Server
from the menu
to automatically search for the TRACE GC Ultra LAN adapter.
3. Double-click on the IP address for the TRACE GC Ultra to configure the adapter.
4. In the Configuration window, click the
OP_mode
tab and select the
Change OP_mode
checkbox. Then click the
More Settings
button.
5. In the TCP Server window, set the Inactivity time field to 30,000 ms and click
OK
.
6. Click
OK
again to upload these settings to the LAN adapter. When the settings have been
uploaded successfully, you can close the NPort utility and proceed to the Instrument
Configuration process for the TRACE GC Ultra.
To configure the Network Enabler
1. Open the Network Enabler application by selecting
Start | All Programs | Network
Enabler
.
2. Under Configuration, select
Broadcast Search
.
Note
The Xcalibur 4.0 Core Components for GC-MS disc also includes the validation
certificate for the Xcalibur core, the release notes for Foundation and Xcalibur and folders
containing the manuals for these products. This content can be explored from the
Xcalibur 4.0 Setup window using the Browse CD and Validation Certificate buttons.
Содержание ISQ 7000
Страница 1: ...ISQ 7000 Mass Spectrometers Hardware Manual 1R120617 0002 Revision C December 2018 ...
Страница 8: ......
Страница 174: ...2 Troubleshooting Contacting Technical Support 146 ISQ 7000 Hardware Manual Thermo Scientific ...
Страница 400: ...3 Advanced Maintenance Replacing Fans and Filters 372 ISQ 7000 Hardware Manual Thermo Scientific ...