3
Advanced Maintenance
Replacing Components of the Analyzer
152
ISQ 7000 Hardware Manual
Thermo Scientific
2. Disconnect the electrometer board cable and large feedthrough cables.
3. Use a T20 Torxhead screwdriver to remove the four screws at each corner of the electron
multiplier plate.
Figure 116.
Disconnecting the Electron Multiplier Cables
Figure 117.
Disconnecting the Electron Multiplier Plate
Electron
Multiplier
Plate
Large
Feedthrough
Cables
Electrometer
Board Cable
Electron Multiplier
Plate Screws
Groove