3
Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
Thermo Scientific
ISQ 7000 Hardware Manual
321
8. Incrementally, and very carefully, disconnect the PC communication board from the
controller interface board by separating the two connectors. Inspect the pins to be certain
they are not bent.
9. Reattach the existing or new PC communication board to the existing or controller
interface board. Be careful not to bend the pins.
10. Inspect the boards from multiple angles to ensure all pins are inserted into the connectors.
It is possible to shift the board and miss a row of pins when you plug the boards together.
11. Reconnect the Ethernet cable to the controller interface board.
CAUTION INSTRUMENT DAMAGE:
Be careful not to exert too much pressure when
separating these boards. Flexing the boards will cause damage.
ATTENTION DOMMAGES À L’INSTRUMENT
: faites attention à ne pas exercer une pression
trop importante lorsque vous séparez ces cartes. Le fait de plier les cartes provoquera des
dommages.
Figure 349.
Replacing the PC Communication Board
Controller Interface Board
PC Communication
Board
Ethernet
Cable
Connector
Connectors