Advanced Maintenance
Replacing the Calibration Gas Components
Thermo Scientific
ISQ 7000 Hardware Manual
355
4. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
5. Open the manifold door and swing it out of the way.
6. Look through the top of the instrument and locate the RF board. It is the second board
standing on end toward the center of the ISQ 7000 instrument. Use a T20 Torxhead
screwdriver to loosen the screw holding the board and bracket to the chassis.
Figure 398.
Opening the Manifold Door
Figure 399.
Removing the RF Board and Bracket
Screws
RF Board
Chassis
Screw
Bracket