1
Performing Routine Maintenance
Replacing a Dual Filament
94
ISQ 7000 Hardware Manual
Thermo Scientific
14. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
15. Open the front door to move the magnet yoke away from the filament.
CAUTION INSTRUMENT DAMAGE:
Use clean, lint-free gloves when touching
components inside the vacuum manifold.
ATTENTION DOMMAGES À L’INSTRUMENT
: utilisez des gants propres non pelucheux
lorsque vous touchez des composants à l’intérieur du collecteur à vide.
Figure 92.
Opening the Manifold Door
Screws