Performing Routine Maintenance
Cleaning the Ion Source Cartridge
76
ISQ 7000 Hardware Manual
Thermo Scientific
10. Press the
Evacuate
button on the front of the instrument. When the
Evacuate
light
flashes green for approximately 20-30 seconds and turns off, the
Ready to Open
light
turns solid green. At that point, the air has been evacuated from the instrument and it is
safe to open it.
11. Pull the vacuum interlock handle up when the
Ready to Open
light is a solid green.
12. Twist the handle of the tool slightly to the left until it is feels like it is lodged into the
left-most track.
13. Push the handle toward the instrument until the end of the handle aligns with the
engraved line at the end of the barrel. When you reach this line, the tool is all the way in
and the ion source cartridge has been placed back onto the ion source block.
IMPORTANT
Make sure you take the correct track on the tool or the ion volume will
disassemble inside the instrument. If that happens, the tool will get stuck and you will
have to shut down, vent the ISQ 7000 instrument, and manually remove the ion volume
through the top of the instrument.
Figure 71.
Pushing the Tool into the Instrument
CAUTION INSTRUMENT DAMAGE:
Make sure the vacuum interlock handle is all the
way up. Otherwise if you insert the tool with the handle down, you can damage the
instrument.
ATTENTION DOMMAGES À L’INSTRUMENT
: assurez-vous que la poignée de verrouillage
du vide est totalement relevée. Dans le cas contraire, si vous insérez l’outil avec la poignée
abaissée, vous risquez d’endommager l’instrument.