Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
322
ISQ 7000 Hardware Manual
Thermo Scientific
12. Install the controller interface board by aligning the two connectors on the bottom of the
board to the connectors on the distribution board beneath it. It will drop into the slot
when it is in the right position.
13. Push the board down into the connectors.
14. Reattach the support bracket.
15. Reattach the top cover and left side panels and tighten the screws.
16. Restart your system by following the instructions in
Figure 350.
Reinstalling the Controller Interface Board
Note
The new PC communication board will not have tune files or detector gain
calibration information stored in memory, so you will need to load a good tune file and
save it to the instrument in manual tune. A factory tune file must also be saved to the
instrument by loading a good default tune file and then pressing the
Alt
key at the same
time as selecting save to instrument. Once the tune file is saved to the PC communication
board memory, a default tune should be run with a clean EI ion source to set the detector
gain calibration.
Connectors
Содержание ISQ 7000