4
Upgrade Equipment
Turbomolecular Pumps
400
ISQ 7000 Hardware Manual
Thermo Scientific
21. Reinstall the power supply and make sure it is flush against the top of the chassis and the
screw holes line up
22. Connect the power cables on the back of the instrument. Make sure you attach the cables
properly. The foreline pump connects to the System Mech Pump connector on the back
of the instrument.
23. Replace the right side panel and attach the screws holding it in place.
24. Slide the instrument closer to the GC.
25. Reattach the top cover panel and tighten the screw holding the top cover panel in place.
26. Restart your system by following the instructions in