3
Advanced Maintenance
Replacing Components of the Manifold
Thermo Scientific
ISQ 7000 Hardware Manual
255
Delivery Tubes
To replace the delivery tubes
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Remove the analyzer tray from the instrument by following the instructions in
“Removing the Analyzer Tray from the ISQ 7000 Instrument”
3. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or the column. If you are using
hydrogen as a carrier gas, power off the GC.
Figure 267.
Opening the Manifold Door
Screws