Reattaching the Analyzer Tray Components
To reattach the analyzer tray components
8. Reattach the ion source to the analyzer tray and reattach the two screws.
1. Reattach the screw holding the lens heater block grounding strap to the analyzer tray.
2. Make sure the ion guide is aligned parallel to the flat on the endcap.
3. Tighten the T10 screws inside the endcap.
4. Slide the ion guide and endcap back into the analyzer tray.
5. Insert the quadrupole assembly with exit endcap onto the analyzer tray.
Figure 130.
Reattaching the Ion Source to the Analyzer Tray
IMPORTANT
When you reassemble the ion source block and lens heater block, ensure
that the blocks are centered properly in the analyzer tray. Check that the ion source
cartridge provided with the instrument can easily slide in and out of the blocks.
Note
Make sure the entrance lens sits flush and centered in the entrance lens endcap.
Figure 131.
Rejoining the Ion Guide to the Quadrupole
Screws
Ion Source
Block
Grounding
Strap
Q1 Quad Assembly
Ion Guide